Department of Chemistry

Heat Exchangers

The chemistry department's helium recovery system at Texas A&M University in College Station, TX makes its way through much of the '72 wing of the department. The helium gas collected during a fill though is extremely cold, and condensation on the tubing could become a real issue. To mitigate this, heat exchangers were put inline immediately after the magnets to raise the fill exhaust to room temperature.

VnmrS 300
Part Description Company Catalog # Quantity
butt-weld NW40 McMaster Carr 4518K42 2
0.5 psi pop-off valve Circle Valve 532T1-.5 2
0.15 psi check valve Circle Valve 259t-1pp-.15 1
4ft exchanger Grainger 5E302 2
2 socket elbow McMaster Carr 5520K64 4
1 socket elbow McMaster Carr 5520K188 2
NW25 ball valve Ideal Vacuum P103569 1
1¼" copper pipe Sourced Local 5.5 ft
spool soft solder Sourced Local 1
flux Sourced Local 1
⅛" to ¼" push-to-connect Grainger 6jk43 4
flow meter Omega FL7611 1
aluminum unistrut 8' McMaster Carr 3230T26 2
unistrut nuts McMaster Carr 3259T32 4
bolts for nuts McMaster Carr 91309A622 4
washers McMaster Carr 98180A130 4
KF25 ¾" NPT-M LDS Vacuum NW25x75NPT 2
KF40/KF25 Reducer LDS Vacuum NW40x25-CR 2
KF25 Cross LDS Vacuum NW25-100-X 2
KF25 ⅛" NPT-F LDS Vacuum NW25-100-SP1 4
1 / 6
2 / 6
3 / 6
4 / 6
5 / 6
6 / 6

Bruker Solids 400 MHz
Waltz Heat Exchanger
Oxford Exhaust Connection
Bruker Exhaust Connection
During a Fill
Flow Meter Plate